Calibration and Verification
Calibrating the 5700A/5720A Series II to External Standards 3
3-7
4. Enter the ambient temperature; then press E. The display shows:
PREV
MENU
IF3-4.EPS
5. Connect the 732B to the calibrator as shown in Figure 3-1.
732B DC STANDARD
AC PWR IN CAL CHARGE LOW BAT
10V 1.018V CHASSIS
GUARD1.018V COM10V COM
SERIAL NUMBER
HI
HI
LO
LO
HI
OUTPUT
V A
SENSE
V
AUX
CURRENT
GUARD
GROUND
WIDEBAND
ΩΩ
CALIBRATOR
VOLTAGE REFERENCE
STANDARD
F7-3.EPS
Figure 3-1. 732B External Calibration Connections.