4
Specifications
Ranges
mA ....................................................... 0 mA to 24 mA
Volts ..................................................... 0 V dc to 30 V dc
Resolution
mA Ranges ........................................... 1 μA
Voltage Range ...................................... 1 mV
Accuracy ..................................................... 0.01 % ±2 LSD all ranges (@23 °C ±5 °C)
Stability ...................................................... 20 ppm of F.S. /°C from -10 °C to 18 °C and
28 °C to 55 °C
Operating Temp Range .............................. -10 °C to 50 °C (14 °F to 122 °F)
Storage Temp Range .................................. -20 °C to 60 °C (-4.0 °F to 140 °F)
Altitude ....................................................... 3000 meters
Ingress Protection Rating ........................... IEC 60529: IP40
Humidity Range ......................................... 10 % to 95 % non-condensing
Display ....................................................... 128 x 64 pixels, LCD Graphic with backlight,
8.6 mm high digits
Power .......................................................... Six AAA alkaline batteries
Battery Life (Alkaline Batteries) ................ ≥40 hours continuous use (measure mode)
Loop Compliance Voltage .......................... 24 V dc @ 20 mA
Loop Drive Capability ................................ 1200 Ω without HART resistor, 950 Ω with
HART resistor
Electromagnetic Environment .................... IEC 61326-1 (portable equipment)
Dimensions (LxWxD) ................................ (152 x 93 x 44) mm, (6.0 x 3.7 x 1.7) in
Weight ........................................................ 0.3 kg (9.5 oz)