mA Loop Calibrator
Required Test Equipment
7
Required Test Equipment
The equipment listed in Table 2 is necessary to do the performance verification tests and
calibration adjustment.
Table 2. Required Equipment for Verification and Calibration
Equipment Recommended Model Minimum Specification
DC Calibrator
Fluke 5522A Multi-Product
Calibrator
DC Voltage: 0 V to 30 V
Accuracy: ±0.002 % +0.5 mV
DC Current 0 mA to 24 mA
Accuracy: ±0.002 % +0.5 μA
Digital Multimeter Fluke 8508A
DC Current: 0 mA to 26 mA
Accuracy: ±0.002 % +0.5 μA
HART Temperature Transmitter Rosemount 644 -
Flexible Test Leads Fluke TL24 -
Double Banana Jack Cable - -
Performance Verification Tests
Warning
To prevent possible electrical shock, fire, or personal injury, do
not perform the performance test or verification test procedures
unless the Product is fully assembled.
The performance tests verify the complete operation of the Product and measure the
accuracy of each function against Product specifications. If the Product fails a part of the
test, calibration adjustment or repair is necessary. See “Calibration Adjustment
Procedure”.
The performance verification tests check the accuracy of each Product function against its
specifications. If the Product fails any of these tests, calibration adjustment or repair is
necessary.