709/709H
Calibration Manual
10
5522A
8508A
hid014.eps
Figure 3. DC Current Measurement Verification Connections
7. Adjust the Calibrator to output the value of test 4 in Table 3 as measured by the DMM,
adjust the Calibrator if necessary to cause the DMM to show the necessary value. Then
verify the display readings on the Product.
8. Repeat for tests 5 and 6.The reading on the Product display should be within the
minimum and maximum values shown in Table 3.
9. Disconnect the test leads from the Calibrator, DMM, and the Product terminals.