EasyManua.ls Logo

Fluke 725Ex - Safety Warnings

Fluke 725Ex
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multifunction Process Calibrator
Specifications
7
Temperature, Thermocouples
Type Range Measure and Source Accuracies
J
-200 to 0 °C
0 to 1200 °C
1.0 °C
0.7 °C
K
-200 to 0 °C
0 to 1370 °C
1.2 °C
0.8 °C
T
-200 to 0 °C
0 to 400 °C
1.2 °C
0.8 °C
E
-200 to 0 °C
0 to 950 °C
0.9 °C
0.7 °C
R
-20 to 0 °C
0 to 500 °C
500 to 1750 °C
2.5 °C
1.8 °C
1.4 °C
S
-20 to 0 °C
0 to 500 °C
500 to 1750 °C
2.5 °C
1.8 °C
1.5 °C
B
600 to 800 °C
800 to 1000 °C
1000 to 1800 °C
2.2 °C
1.8 °C
1.4 °C
L
-200 to 0 °C
0 to 900 °C
0.85 °C
0.7 °C
U
-200 to 0 °C
0 to 400 °C
1.1 °C
0.75 °C
N
-200 to 0 °C
0 to 1300 °C
1.5 °C
0.9 °C
XK
-200 to 100 °C
-100 to 800 °C
0.5 °C
0.6 °C
BP
0 to 800 °C
800 to 2500 °C
1.2 °C
2.5 °C
Resolution:
0.1 °C, 0.1 °F
Loop Power Supply
Voltage: 12 V
Maximum current: 24 mA
Short circuit protected
Entity Parameters
For Entity Parameters, Refer to Fluke 725Ex CCD for use in Ex-hazardous areas.
RTD Excitation (simulation)
Allowable Excitation by RTD type
Ni 120 0.15 to 3.0 mA
Pt 100-385 0.15 to 3.0 mA
Pt 100-3926 0.15 to 3.0 mA
Pt 100-3916 0.15 to 3.0 mA
Pt 200-385 0.05 to 0.80 mA
Pt 500-385 0.05 to 0.80 mA
Pt 1000-385 0.05 to 0.40 mA

Table of Contents

Other manuals for Fluke 725Ex

Related product manuals