Digital Communication Interface
Cooling Control10
10-7
Table 10-1. Interface Command Summary (cont.)
Command
Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Ramp and Soak Menu
Read number of
programmable
set-points
pn pn pn: 9 pn: 2
Set number of
programmable
set-points to n
pn=n pn=4
Read programmable set-
point
number n
psn ps3 psn: 9999.99
{C
or F}
ps1: 50.00
C
Set programmable set-
point number
n to n
psn=n ps3=50 1 to 8,
Instrument
Range
40 to 150°C
Read program set-point
soak time
pt
pt
ti: 999 ti: 5
Set program set-point
soak time
to n minutes
pt=n pt=5 0 to 500
Read program control
mode
pc pc prog: {OFF or
ON}
prog: OFF
Set program control
mode:
pc=g[o]/s[top]/c
[ont]
GO or STOP
or CONT
Start program pc=g[o] pc=g
Stop program pc=s[top] pc=s
Continue program pc=c[ont] pc=c
Read program function pf pf pf: 9 pf: 3
Set program function to n pf=n pf=2 1 to 4
Secondary Menu
Read proportional band
setting
pr[op-band] pr pr: 9.999 pr: 0.326
Set proportional band to
n
pr[op-band]=n pr=0.326
Read cutout setting c[utout] c cu: 999
{x},{xxx}
cu: 160 C,
in
Set cutout setting: c[utout]=n/r[ese
t]
Set cutout to n degrees c[utout]=n c=160 Temperature
Range
Reset cutout now c[utout]=r[eset] c=r
Read heater power (duty
cycle)
po[wer] po po: 9999 po: 1