7341
Users Guide
10-8
Table 10-1. Interface Command Summary (cont.)
Command
Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Configuration Menu
Probe Menu
Read R0 calibration
parameter
r[0] r r0: 999.999 r0: 100.578
Set R0 calibration
parameter to n
r[0]=n r=100.324 98.0 to
104.999
Read ALPHA calibration
parameter
al[pha] al al: 9.9999999 al:
0.0038573
Set ALPHA calibration
parameter
to n
al[pha]=n al=0.0038433 .00370 to
.0039999
Operating Parameters
Menu
Read cutout mode cm[ode] cm cm: {xxxx} cm: auto
Set cutout mode: cm[ode]=r[eset]
/a[uto]
Set cutout to be reset
manually-
cm[ode]=r[eset] cm=r
Set cutout to be reset
automatically
cm[ode]=a[uto] cm=a
Serial Interface Menu
Read serial sample setting
sa[mple] sa sa: 9999 sa: 1
Set serial sampling
setting to n seconds
sa[mple]=n sa=0 0 to 4000
Set serial duplex mode:
du[plex]=f[ull]/h
[alf]
FULL or
HALF
Set serial duplex mode to
full
du[plex]=f[ull] du=f
Set serial duplex mode to
half
du[plex]=h[alf] du=h
Set serial linefeed mode:
lf[eed]=on/of[f]
Set serial linefeed mode
to on
lf[eed]=on lf=on
Set serial linefeed mode
to off
lf[eed]=of[f] lf=of
Calibration Menu
Read C0 calibration
parameter
*c0 *c0 c0: 9.9999 c0: 0.0002
Set C0 calibration
parameter to n
*c0=n *c0=0 Unlimited
Read CG calibration
parameter
*cg *cg cg: 999.99 cg: 406.25