Documenting Process Calibrator
Specifications
105
DC Voltage Measurement
Range Resolution % of Reading +% of Full Scale
1 Year 2 Year
110 mV 1 µV 0.025% + 0.015% 0.05% + 0.015%
1.1 V 10 µV 0.025% + 0.005% 0.05% + 0.005%
11 V 100 µV 0.025% + 0.005% 0.05% + 0.005%
110 V 1 mV 0.05% + 0.005% 0.1% + 0.005%
300 V 10 mV 0.05% + 0.005% 0.1% + 0.005%
Temperature Coefficient: (0.001% of rdg. + 0.0015% f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C
Input Impedance: 5 M
Common Mode Error: 0.008% f.s./(Common Mode Volt)
Maximum Input Voltage: 300 V rms