EasyManua.ls Logo

Fluke 744 - DC Current Measurement; Resistance Measurement

Fluke 744
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Documenting Process Calibrator
Specifications
107
DC Current Measurement
Range Resolution % of Reading +% of Full Scale
1 Year 2 Year
30 mA 1 µA 0.01% + 0.015% 0.02% + 0.015%
110 mA 10 µA 0.01% + 0.015% 0.02% + 0.015%
Temperature Coefficient: (0.001% of rdg. + 0.002% f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.01% f.s./(Common Mode Volt)
Maximum Input Voltage: 30 V dc
Resistance Measurement
Range Resolution % of Reading + ohms
1 Year 2 Year
11 0.001 0.05% + 0.05 0.075% + 0.05
110 0.01 0.05% + 0.05 0.075% + 0.05
1.1 k 0.1 0.05% + 0.5 0.075% + 0.5
11 k 1 0.1% + 10 0.1% + 10
Temperature Coefficient: (0.01% f.s. + 2 m Ω) /°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.005% f.s/(Common Mode Volt)
Maximum Input Voltage: 30 V dc

Table of Contents