Simulating RTDs
Note
Refer to Table 6 for information about RTD
(Resistance-Temperature Detector) types
supported by the calibrator.
Connect the calibrator to the instrument under test as
shown in Figure 21. The figure shows connections
for two, three, or four-wire transmitters. For three or
four-wire transmitters, use the 4-inch long stackable
jumper cables to connect the third and fourth wires at
the source V RTD jacks.
Proceed as follows to simulate an RTD (Resistance-
Temperature Detector):
1. If necessary, press M for SOURCE mode.
2. Press t until the select RTD type display is
showing.
3. Press the u or d keys followed by e to
select the desired RTD type.
4. Enter the temperature you want to simulate as
prompted by the display, then press e.