Fluke 744 - Measurement Scale; Linear-Output Transmitters

Fluke 744
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Documenting Process Calibrator
Using Measure Mode
41
Measurement Scale
This feature lets you scale the measurements in
accordance with a particular process instrument’s
response. Percent of scale works for linear-output
transmitters or square-law transmitters such as
differential pressure transmitters that report flow rate.
Linear-Output Transmitters
1.
If necessary, press M for MEASURE mode.
2.
Select a measurement function (m,v,
h,q, t, or p) as previously described.
3.
Press the
Scale softkey.
4.
Select
% scale from the list.
5.
Use the numeric keypad to enter the 0% of scale
value (
0% Value).
6.
Press e.
7.
Use the numeric keypad to enter the 100% of
scale value (
100% Value).
8.
Press e.
9.
Press the
Done softkey.
Percent of scale remains in effect until you change to
another measurement function or press the
Scale
softkey and select another scale mode.