Documenting Process Calibrator
Calibrating a Process Instrument
69
8. Press M for MEASURE/SOURCE mode. The
display changes to:
gj42s.eps
9. Press the As Found softkey, followed by the
Instrument softkey. The display changes to:
gj44s.eps
10. Enter values for 0% and 100% of 4.0 mA and
20.0 mA, respectively. Set
Tolerance to 0.5% of
span. (Use other values if necessary for your
application.)