ProcessMeter 
  Measuring Electrical Parameters
   
 19 
6.  The diode is good if it passes the tests in steps 4 
and 5. 
Displaying Minimum, Maximum, and Average 
MIN MAX recording stores the lowest and highest 
measurements, and maintains the average of all 
measurements. 
Press M to turn on MIN MAX recording. Readings are 
stored until the meter is turned off, switched to another 
measurement or source function, or MIN MAX is turned 
off. The beeper sounds when a new maximum or 
minimum is recorded. Auto power-off is disabled and auto 
ranging is turned off during MIN MAX recording.  
Press M again to cycle through the MAX, MIN, and 
AVG displays. Press and hold M for 1 second to erase 
stored measurements and exit. 
If MIN MAX recording is on continuously for over 40 hours, 
minimum and maximum readings are still recorded, but 
the displayed average no longer changes. 
In MIN MAX recording, press H to suspend recording; 
press H again to resume recording.  
Using AutoHold 
Note 
MIN MAX recording must be off to use AutoHold. 
! Warning 
To avoid possible electric shock, do not use 
AutoHold to determine if dangerous voltage 
is present. AutoHold will not capture 
unstable or noisy readings. 
Activate AutoHold to freeze the meter's display on each 
new stable reading (except in the frequency counter 
mode). Press H to activate AutoHold. This feature 
allows measurements to be taken in situations in which it 
is difficult to look at the display. The meter beeps and 
updates the display with each new stable reading.