ProcessMeter
Specifications
45
Accuracy adders for use in RF Fields: In an RF field of
3 V/m, change the accuracy specifications as follows:
For AC Volts Measurement, add 0.25 % of range
For DC Current Measurement, 30.000 mA range, add
0.14 % or range
For DC Current Output, add 0.32 % of span
Accuracy for all meter functions is not specified in RF’
fields > 3 V/m.
Relative humidity: 95 % up to 30 °C, 75 % up to 40 °C,
45 % up to 50 °C, and 35 % up to 55 °C
Vibration: Random 2g, 5 to 500 Hz
Shock: 1 meter drop test
Safety: The Meter complies with, EN61010-1:2001,
ANSI/ISA S82.01-2004, CAN/CSA C22.2 No. 1010-
1:2004, UL61010-1, Measurement Category III, 1000V,
Pollution Degree 2 and Measurement Category IV, 600V,
Pollution Degree 2.
Certifications:
), !To 61010-1 2nd Edition,
®, ;
Power requirements: Four AA batteries (alkaline
recommended)
Size: 10.0 cm X 20.3 cm X 5.0 cm (3.94 in X 8.00 in X
1.97 in)
Weight: 610 g (1.6 lbs)