80 Series III
Service Information
6
Current Specifications
Accuracy
1
Function Range Resolution
Model 83
2
Model 85
3,4
Model 87
3,4
Burden
Voltage
(typical)
mA \
(45 Hz to
2 kHz)
40.00 mA
400.0 mA
4000 mA
10.00 A
5
0.01 mA
0.1 mA
1 mA
0.01 A
±(1.2 % + 2)
6
±(1.2 % + 2)
6
±(1.2 % + 2)
6
±(1.2 % + 2)
6
±(1.0 % + 2)
6
±(1.0 % + 2)
6
±(1.0 % + 2)
6
±(1.0 % + 2)
6
±(1.0 % + 2)
±(1.0 % + 2)
±(1.0 % + 2)
±(1.0 % + 2)
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
mA [ 40.00 mA
400.0 mA
4000 mA
10.00 A
5
0.01 mA
0.1 mA
1 mA
0.01 A
±(0.4 % + 4)
±(0.4 % + 2)
±(0.4 % + 4)
±(0.4 % + 2)
±(0.2 % + 4)
±(0.2 % + 2)
±(0.2 % + 4)
±(0.2 % + 2)
±(0.2 % + 4)
±(0.2 % + 2)
±(0.2 % + 4)
±(0.2 % + 2)
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
µ
A B
(45 Hz to
2 kHz)
400.0 µA
4000 µA
0.1 µA
1 µA
±(1.2 % + 2)
6
±(1.2 % + 2)
6
±(1.0 % + 2)
6
±(1.0 % + 2)
6
±(1.0 % + 2)
±(1.0 % + 2)
100 µV/µA
100 µV/µA
µ
AF
400.0 µA
4000 µA
0.1 µA
1 µA
±(0.4 % + 4)
±(0.4 % + 2)
±(0.2 % + 4)
±(0.2 % + 2)
±(0.2 % + 4)
±(0.2 % + 2)
100 µV/µA
100 µV/µA
1. For extended use at high RH, change the 400 mV and 4.0 V ac accuracy specification to ± (1.5 % + 4 counts) for
45 Hz to 1 kHz and ± (3.0 % + 25 counts) for 5 kHz to 20 kHz.
2. AC conversion for Model 83 is ac coupled and calibrated to the rms value of a sinewave input.
3. AC conversions for Models 85 and 87 are ac coupled, true rms responding, and valid from 5 % to 100 % of
range.
4. See note 2 in under “
AC Voltage Specifications (Models 85 and 87).”
5. 10 A continuous; 20 A for 30 seconds maximum: Accuracy unspecified over 10 A.
6. Below a reading of 200 counts, add 10 counts.
Capacitance and Diode Function Specifications
Function Range Resolution Accuracy
1
E 5.00 nF
0.0500 µF
0.500 µF
5.00 µF
0.01 nF
0.0001 µF
0.001 µF
0.01 µF
±(1 % + 3)
±(1 % + 3)
±(1 % + 3)
±(1.9 % + 3)
G 3.000 V 0.001 V ±(2 % + 1)
1. With a film capacitor or better, using Relative mode to zero residual. See “Accuracy” earlier in the specifications
for a complete explanation.
Frequency Counter Specifications
Function Range Resolution Accuracy
1
Frequency
(0.5 Hz to 200 kHz,
pulse width >2 µs)
199.99
1999.9
19.999 kHz
199.99 kHz
>200 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
±(0.005 % + 1)
±(0.005 % + 1)
±(0.005 % + 1)
±(0.005 % + 1)
unspecified
1. See “Accuracy” earlier in the specifications for a complete explanation.

Other manuals for Fluke 80 Series III

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 80 Series III and is the answer not in the manual?

Fluke 80 Series III Specifications

General IconGeneral
BrandFluke
Model80 Series III
CategoryMultimeter
LanguageEnglish

Related product manuals