80i-110s Instruction Manual 7
ELECTRICAL SPECIFICATIONS
All Electrical Specifications are valid at a temperature of 23
q
C
r
3
q
C (73
q
F
r
5
q
F).
Current Ranges:
0 to 10A dc or ac peak
0 to 100A dc or ac peak
Output Signals:
10A range: 100 mV/A
100A range: 10 mV/A
Working voltage (Clamps jaws to Ground):
600V ac rms on Installation Category II per IEC 1010-1 circuits.
300V ac rms on Installation Category III per IEC 1010-1 circuits.
Floating Voltage (Output cable and connector to Ground):
600V ac rms on Installation Category II per IEC 1010-1 circuits.
300V ac rms on Installation Category III per IEC 1010-1 circuits.
Basic Accuracy (DC to 1kHz):
Extended Accuracy:
For other frequencies, refer to the appropriate input current range and add the
error listed below to the "Basic Accuracy" error.
Input Current
(DC or AC peak)
Error (after zero check)
100 mV/A 10 mV/A
0 to 10A
0 to 40A
40 to 80A
80 to 100A
<3% of reading +50 mA
-
-
-
-
<4% of reading +50 mA
<12% of reading +50 mA
<15% of reading
Frequency
Additional Error
100 mV/A 10 mV/A
1 to 5 kHz
5 to 20 kHz
>20 kHz
3%
12%
not specified
3%
12%
not specified