Specifications
General Specifications
8
8-3
Specifications:
The following specifications apply within 1 year of calibration when operating
in a temperature range of 18°C (64°F) to 28°C (82°F).
AC Volts (True RMS, AC-Coupled) [
±
(percent of reading + counts)]
Range
Frequency
20-50 Hz 50-100 Hz 1 kHz-30 kHz 30 kHz-100 kHz 100kHz-200kHz 200 kHz-300 kHz
300.00
mV
1.5% + 10
0.19 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
4% + 200
1.39 dB
8% + 200
1.68 dB
10%+ 200
1.82 dB
3.0000V 1.5% + 10
0.19 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
4% + 200
1.39 dB
8% + 200
1.68 dB
10%+ 200
1.82 dB
30.000V 1.5% + 10
0.19 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
4% + 200
1.39 dB
8% + 200
1.68 dB
10%+ 200
1.82 dB
300.00V 1.5% + 10
0.19 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
4% + 200
1.39 dB
8% + 200
1.68 dB
10%+ 200
1.82 dB
1000.0V 1.5% + 10
0.19 dB
1.5% + 10
0.19 dB
NA NA NA NA
Crest Factor:
300 mV - 300V ranges 3:1, 1000V range 3:1 decreasing to 1.41:1 as input voltage increases to
1000V (peak voltage not to exceed 1414V).
Measurement Range:
300 mV range from 10% to 100% of range.
3V - 1000V ranges from 5% to 100% of range.
For frequencies > 100 kHz 30% to 100% of range (all ranges).