EasyManua.ls Logo

Fluke 87V Ex - Ex-Certification Data

Fluke 87V Ex
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
87V EX
Calibration Manual
6
Detailed Specifications
For all detailed specifications:
Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18 ºC to 28 ºC, with relative humidity up to
90 %, for a period of one year after calibration.
In the 4 ½-digit mode, multiply the number of least significant digits (counts) by 10. AC conversions are ac-coupled and
valid from 3 % to 100 % of range. The Meter is true rms responding. AC crest factor can be up to 3 at full scale, 6 at half
scale. For non-sinusoidal wave forms add -(2 % Rdg + 2 % full scale) typical, for a crest factor up to 3.
AC Voltage Function
Function Range Resolution Accuracy
45 - 65 Hz 30 - 200 Hz 200 - 440 Hz 440 Hz - 1 kHz 1 - 5 kHz 5 - 20 kHz
[1]
± (0.7 % + 4)
± (0.7 % + 2)
± (2.0 % + 4)
± (2.0 % + 20)
600.0 m
V
6.000 V
60.00 V
600.0 V
0.1 mV
0.001 V
0.01 V
0.1 V
± (2.0 % + 4)
[3]
unspecified
1000 V 1 V
± (1.0 % + 4)
unspecified unspecified
K
[2,4]
Low pass filter ± (0.7 % + 2) ± (1.0 % + 4) +1 % + 4
-6 % - 4
[5]
unspecified unspecified unspecified
[1] Below 10 % of range, add 12 counts.
[2] The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a
residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2-digit change for readings over 3 % of range.
Using REL to offset this reading may produce a much larger constant error in later measurements.
[3] Frequency range: 1 kHz to 2.5 kHz.
[4] A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.
[5] Specification increases from -1% at 200 Hz to -6% at 440 Hz when filter is in use.
DC Voltage, Resistance, and Conductance Function
Function Range Resolution Accuracy
L
6.000 V
60.00 V
600.0 V
1000 V
0.001 V
0.01 V
0.1 V
1 V
± (0.05 % + 1)
± (0.05 % + 1)
± (0.05 % + 1)
± (0.05 % + 1)
F
mV
600.0 mV 0.1 mV ± (0.1 % + 1)
e
nS
600.0
6.000 k
60.00 k
600.0 k
6.000 M
50.00 M
60.00 nS
0.1
0.001 k
0.01 k
0.1 k
0.001 M
0.01 M
0.01 nS
± (0.2 % + 2)
[1]
± (0.2 % + 1)
± (0.2 % + 1)
± (0.6 % + 1)
± (0.6 % + 1)
± (1.0 % + 3)
[2]
± (1.0 % + 10)
[1,2]
[1] When using the REL function to compensate for offsets.
[2] Add 0.5 % of reading when measuring above 30 M in the 50 M range, and 20 counts below 33 nS in the 60 nS range.
Temperature
Temperature Resolution Accuracy
[1,2]
-200 ºC to +1090 ºC
-328 ºF to +1994 ºF
0.1 ºC
0.1 ºF
1 % + 10
1 % + 18
[1] Does not include error of the thermocouple probe.
[2] Accuracy specification assumes ambient temperature stable to ± 1 ºC. For ambient temperature changes of ± 5 ºC, rated accuracy
applies after 1 hour.

Other manuals for Fluke 87V Ex

Related product manuals