True-rms Multimeter
Detailed Specifications
7
Current Function
Function Range Resolution Accuracy
[1, 2]
B
urden Voltage (typical
mA
\
(45 Hz to 2 kHz)
mA
[
60.00 mA
400.0 mA
[4]
6.000 A
10.00 A
[3]
60.00 mA
400.0 mA
[4]
6.000 A
10.00 A
[3]
0.01 mA
0.1 mA
0.001 A
0.01 A
0.01 mA
0.1 mA
0.001 A
0.01 A
±(1.0 % + 2)
± (1.0 % + 2)
± (1.0 % + 2)
± (1.0 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
µA B
(45 Hz to 2 kHz)
µAF
600.0 µA
6000 µA
600.0 µA
6000 µA
0.1 µA
1 µA
0.1 µA
1 µA
± (1.0 % + 2)
± (1.0 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
100 µV/µA
100 µV/µA
100 µV/µA
100 µV/µA
[1] AC conversions are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except 400 mA range (5 % to 100 %
of range) and 10 A range (15 % to 100 % or range).
[2] The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a
residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2 digit change for readings over 3 % of
range. Using REL to offset this reading may produce a much larger constant error in later measurements
[3] W 10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum; > 10 A
unspecified.
[4] 400 mA continuous; 600 mA for 18 hrs maximum.
Capacitance and Diode Function
Function Range Resolution Accuracy
E
10.00 nF
100.0 nF
1.000 µF
10.00 µF
100.0 µF
9999 µF
0.01 nF
0. 1 nF
0.001 µF
0.01 µF
0.1 µF
1 µF
± (1 % + 2)
[1]
± (1 % + 2)
[1]
± (1 % + 2)
± (1 % + 2)
± (1 % + 2)
± (1 % + 2)
G
3.000 V 0.001 V ± (2 % + 1)
[1] With a film capacitor or better, using Relative mode to zero residual.
Frequency Counter
Function Range Resolution Accuracy
Frequency
(0.5 Hz to 200 kHz,
pulse width > 2 µs)
199.99
1999.9
19.999 kHz
199.99 kHz
> 200 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
± (0.005 % + 1)
± (0.005 % + 1)
± (0.005 % + 1)
± (0.005 % + 1)
unspecified
Frequency Counter Sensitivity and Trigger Levels
Minimum Sensitivity (RMS Sine wave) Approximate Trigger Level
Input Range
[1]
5 Hz - 20 kHz 0.5 Hz - 200 kHz (DC Voltage Function)
600 mV dc
600 mV ac
6 V
60 V
600 V
1000 V
70 mV (to 400 Hz)
150 mV
0.3 V
3 V
30 V
100 V
70 mV (to 400 Hz)
150 mV
0.7 V
7 V (140 kHz)
70 V (14.0 kHz)
200 V (1.4 kHz)
40 mV
1.7 V
4 V
40 V
100 V
Duty Cycle Range Accuracy
0.0 to 99.9 % Within ± (0.2% per kHz + 0.1 %) for rise times < 1 µs.
[1] Maximum input for specified accuracy = 10X Range or 1000 V.