87V EX
Calibration Manual
10
F1
F2
1
aom12f.eps
Figure 1. Opening the Meter, Battery and Fuse Replacement
Reassembling the Meter Case
To reassemble the Meter case:
1. Verify that the rotary knob and circuit board switch are in the OFF position, and that
the gasket remains secured to the bottom case.
2. Place the PCA into the bottom case.
3. Place the case top on the case bottom.
4. To avoid damaging the battery wire, ensure the wire exits the middle of the battery
compartment.
5. Properly seat the case gasket and snap the case halves together above the LCD end.
See Figure 1.
6. Reinstall the three case screws and the battery door.
7. Secure the battery door by turning the screw 1/4-turn clockwise.
8. Go to “Performance Tests” later in this document, and perform the procedures
described.