87V EX
Calibration Manual
22
Table 7. Calibration Adjustment Steps (Cont.)
Function
(Switch Position)
Input Terminal Adjustment Step Input Value
C-24 60.00 mA, 60 Hz
C-25 400.0 mA, 60 Hz
C-26 60.00 mA dc
mA
(Milliamps)
C-27 400.0 mA dc
C-28 600.0 µA ac, 60 Hz
C-29 6000 µA, 60 Hz
C-30 600.0 µA dc
µA
(Microamps)
mA /UA
C-31 6000 µA dc
Service and Parts
Replacement parts are shown in Table 7 and Figure 4. To order parts and accessories,
refer to “Contacting Fluke”.
Table 9. 87V EX Final Assembly
Item Description Qty.
Fluke Part or
Model Number
BT1 W Battery, 9 V (see Table 2) 1
822270
F1 W
Fuse, 0.44 A, 1000 V, FAST 1 943121
F2 W Fuse, 11 A, 1000 V, FAST 1 803293
H1-3 Screw, Case 3 832246
MP1 Shock absorber 1 828541
MP2-3 Foot, Non-Skid 2 824466
MP4 Ex-Holster (Red) 1 2520563
MP5 Battery Door 1 2520595
AC72 Alligator Clip, Black 1 1670652
AC72 Alligator Clip, Red 1 1670641
TL75 Test Lead Set 1 855742
MP6-7 Access Door Fastener 2 948609
80BK Thermocouple Assembly, K-Type, Beaded, Molded Dual Banana
Plug, Coiled
1 1273113
NA Tiltstand of the Ex holster 1 2520056
TM1 87V Ex Users Manual (English, French, & German) 1 2518115
TM2 CD ROM,87V Ex Users Manual 1 2520777
WTo ensure safety, use exact replacement only.