8845A/8846A 
Programmers Manual 
40 
  [3] For period measurements, the Meter uses one range for all inputs 
between 0.33 seconds, and 3.3 µsec. With no input signal applied, 
period measurements return zero. 
  [4] The range and resolution are fixed for temperature measurements. 
However, placing PT100_385 or PT100_392 into the range field will 
set the RTD type. Placing CUST1 into the range field will cause the 
Meter to use the values set into the R
0
 and Alpha parameters. 
  [5] The range and resolution are fixed for diode test: 1 Vdc range and 5½ 
digits. When set to ON, the low current parameter sets the diode test 
current to 0.1 mA. When OFF, the current is 1 mA. When set to ON, 
the high voltage parameter sets the diode test voltage to 10 volts. When 
OFF, the voltage is 5 volts. To specify a diode test voltage, you must 
also include ON or OFF for the low current parameter. 
  [6] The range and resolution are fixed for continuity tests: 1 kΩ range and 
5½ digits. 
Table 18. Preset Conditions for the MEASure? and CONFigure Commands 
Parameter Setting 
AC Filter (DET:BAND)  20 Hz 
Autozero (ZERO:AUTO)  OFF if NPLC <1, ON if NPLC >=1 
DC Digital Filter  ON 
Input Impedance (INP:IMP:AUTO)  OFF 
Samples per trigger (SAMP:COUN)  1 
Trigger count (TRIG:COUN)  1 
Trigger delay (TRIG:DEL)  Automatic 
Trigger source (TRIG:SOUR)  Internal (immediate) 
Math Function (CALCulate subsystem)  OFF 
 
Setting Range and Resolution Parameters 
The MEASure? and CONFigure commands incorporate parameters for setting the 
Meter’s function, its range, and its resolution.  Placing the input signal’s expected 
amplitude into the range value causes the Meter to select a range appropriate for the 
supplied signal amplitude.  
Although the frequency and period functions use a single range for all inputs between 3 
Hz and 300 kHz, it is necessary to specify a range whenever specifing a resolution setting 
for these functions. When specifying resolution, specify it in the units of the function 
instead of number of digits. For ac or dc volts, specify the resolution in volts. The 
resolution specification for frequency is in hertz. 
Note 