85
Appendix B
Fluke 45 Emulation Limitations
Introduction
Due to differences in their design, the 8845A and 8846A (collectively the 884XA) cannot
emulate all the functions and features of the Fluke 45. Some functions are compatible but
are not implemented the same in the 884XA meters.
This appendix covers the differences with, a list of commands that are not compatible,
and information on those functions that will respond differently by the 884XA.
Incompatible Fluke 45 Commands
Due to their design differences, there are Fluke 45 function and modifier commands that
the 884XA meters will not execute. Whenever a non-compatible command is detected by
the 884XA meters, they will generate a -102, syntax error. Table 33 lists Fluke 45
commands that are not compatible with the 8845A and 8846A.
Table 33. Non-compatible Fluke 45 Remote Commands
Command Reason
AACDC The 884XA meters do not have an ac plus dc rms current function.
HOLD The 884XA meters do not have a Touch Hold function.
HOLDCLR The 884XA meters do not have a Touch Hold function.
HOLDTHRESH The 884XA meters do not have a Touch Hold function.
HOLDTHRESH? The 884XA meters do not have a Touch Hold function.
MIN See the section on “Math Operations” for an appropriate replacement.
MAX See the section on "Math Operations" for an appropriate replacement.
MAXSET See the section on "Math Operations" for an appropriate replacement.
MINSET See the section on "Math Operations" for an appropriate replacement.
MINCLR See the section on "Math Operations" for an appropriate replacement.
VACDC The 884XA meters do not have an ac plus dc rms voltage function.