Digital Multimeter 
 
Using FlukeView Forms 
  75 
External Triggering 
A measurement can be externally triggered in three ways: 
 
•  External trigger. 
•  GET command (IEEE Interface only) 
•  *TRG command 
To trigger a measurement over the RS-232 or IEEE 488 computer interface, send the 
Meter a *TRG command over the computer interface. The GET command is only valid 
for the IEEE interface. 
When set for external trigger, the Meter will trigger measurements when a negative-going 
signal is applied to the rear-panel trigger jack. 
Note 
The Fluke 45 provides a method of triggering through its RS-232 port. The 
8845A and 8846A do not provide this method of triggering. 
Table 29. Trigger Types 
Type  Trigger  Rear Trigger  Settling Delay 
1 
2 
3 
4 
5 
Internal 
External 
External 
External 
External 
Enabled
[1] 
Enabled
[1] 
Enabled
[1] 
Enabled 
Enabled 
- 
Off 
On 
Off 
On 
[1]  Although the Fluke 45 disabled the rear-panel trigger for these trigger types, the 884XA leaves the rear-panel 
trigger enabled.
 
Using FlukeView Forms 
You can use FlukeView Forms to download and store readings from the Meter to a PC 
running MicroSoft’s Windows operating system. FlukeView Forms Basic Documenting 
Software must first be installed on the PC. Refer to the FlukeView Forms Installation 
Guide to install FlukeView Forms on your Windows PC. 
Note 
The Meter’s RS-232 port must be enabled and properly configured before 
FlukeView forms will communicate with the Meter. In addition, the 
Fluke 45 command set must be selected in the Meter. 
To connect and configure the Meter to the PC with the installed FlukeView Forms Basic 
Documenting Software: 
1.  Connect a null-modem cable between the RS-232 port of the Meter and the PC’s RS-
232 port. 
2.  Open FlukeView Forms Documenting Software. 
3.  The current COM serial port setting is displayed at the bottom right of the FlukeView 
Forms Basic window. Double-click on it to change the COM port setting to the 
virtual COM port used by the null-modem cable. 
4.  Turn the Meter on. 
5.  Select the Fluke 45 Command set by following the instructions in the Selecting the 
Program Language section of Chapter 3 of the Users Manual.