Digital Multimeter 
 
Supported SCPI Commands 
  61 
*ESE <enable value> 
Description  Enables or disables bits in the Meter’s Standard Event enable register. 
Parameters  <n>  =  0 to 255  Binary-weighted decimal value representing the 
bits to set in the register. Although this register 
has 8 bits and will accept a number from 0 to 
255, only bits 0, 2 through 5, and 7 are used.  
Examples  *ESE 160  Sets bits 5 and 7 to one and all other bits to zero. 
*ESE? 
Description  Retrieves the value in the Meter’s Standard Event enable register. 
Examples  *ESE?  Retrieves a binary-weighted decimal value 
representing the bits set in the Standard Event 
enable register. 
*ESR? 
Description  Retrieves the value in the Meter’s Standard Event register. 
Examples  *ESR?  Retrieves a binary-weighted decimal value 
representing the bits set in the Standard Event 
register. 
*OPC 
Description  Sets bit 0 (Operation complete) in the Meter’s Standard Event register. 
Examples  *OPC  Sets bit 0 in Standard Event register. 
*OPC? 
Description  Returns a one to the Meter’s output buffer. 
Examples *OPC? 
*PSC <n> 
Description  Enables or disables the clearing of the Status Byte Enable register and 
Standard Event Enable register when Meter power is turned on. This 
setting is stored in non-volatile memory. 
Parameters  <n>  =  0  Disables clearing Status Byte and Standard 
Event  Enable register at power-up. 
      1  Enables clearing Status Byte and Standard Event 
Enable register at power-up.