8845A/8846A 
Programmers Manual 
80 
Table 30. Execution Error Codes and Descriptions (cont.) 
Error Code  Error Text  Description 
200  IG Invalid  Invalid response type from inguard. 
201  IG Packet size  Incorrect packet size from inguard. 
202  Trigger ignored  Trigger ignored. 
203 Configuration 
acknowledgement 
Wrong number configuration acknowledgement. 
222 Parameter invalid Invalid parameter. 
223  FUNC2  Function/2nd func mismatch 
224 Math mismatch Function/math mismatch 
225  Range mismatch  Function/range mismatch. 
226 Load fail  Measurement configuration load failed. 
300  GX Sync send  Inguard not responding (send). 
301  GX Sync recv  Inguard not responding (recv), 
302  GX Sync error  Lost sync with inguard. 
303  GX Port fail  Could not open guard crossing port. 
310  GX Ack queue full  Acknowledgement queue full. 
311  GX Pack sequence  Acknowledgement queue full. 
320  GX Ack  Guard crossing not acking packets. 
321  GX Pack over  Guard crossing oversize packet recieved. 
322  GX Ack num  Guard crossing wrong ack number. 
325  GX Ack byte  Guard crossing unknown ack byte. 
330  GX CRC failed  Guard crossing  Bad CRC. 
340  GX Info packet  Guard crossing info packet recieved; link not active. 
350  GX Control byte  Guard crossing unknown control byte. 
360  GX Quality low  Guard crossing  quality indicator too low. 
370  GX Restart  Guard crossing restarted. 
421 Instrument configuration 
inv 
Tried to set invalid state. 
422 Instrument configuration 
parameter 
Invalid parameter. 
426 Instrument configuration 
load 
Instrument configuration load failed. 
510  CCO Name bad  CCO constant name is bad. 
511 RS-232 
framing/parity/overrun 
error