Table 30. Execution Error Codes and Descriptions (cont.)
Error Code Error Text Description
200 IG Invalid Invalid response type from inguard.
201 IG Packet size Incorrect packet size from inguard.
202 Trigger ignored Trigger ignored.
203 Configuration
acknowledgement
Wrong number configuration acknowledgement.
222 Parameter invalid Invalid parameter.
223 FUNC2 Function/2nd func mismatch
224 Math mismatch Function/math mismatch
225 Range mismatch Function/range mismatch.
226 Load fail Measurement configuration load failed.
300 GX Sync send Inguard not responding (send).
301 GX Sync recv Inguard not responding (recv),
302 GX Sync error Lost sync with inguard.
303 GX Port fail Could not open guard crossing port.
310 GX Ack queue full Acknowledgement queue full.
311 GX Pack sequence Acknowledgement queue full.
320 GX Ack Guard crossing not acking packets.
321 GX Pack over Guard crossing oversize packet recieved.
322 GX Ack num Guard crossing wrong ack number.
325 GX Ack byte Guard crossing unknown ack byte.
330 GX CRC failed Guard crossing Bad CRC.
340 GX Info packet Guard crossing info packet recieved; link not active.
350 GX Control byte Guard crossing unknown control byte.
360 GX Quality low Guard crossing quality indicator too low.
370 GX Restart Guard crossing restarted.
421 Instrument configuration
inv
Tried to set invalid state.
422 Instrument configuration
parameter
Invalid parameter.
426 Instrument configuration
load
Instrument configuration load failed.
510 CCO Name bad CCO constant name is bad.
511 RS-232