EasyManua.ls Logo

Fluke 893A - Versus Range Checks; Null Detector

Fluke 893A
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
n. Connect
the
output of the
dc voltage calibrator
between
the
LO and
HI INPUT terminals.
o. Apply
the dc
input
voltages versus the
ranges in-
dicated in
Figure
4-5, observing that the meter
indicates
full-scale
(+1)
± 1
-
1/2
small divisions.
4-30.
When
the results
of the
checks agree with the
information
given, the
Null
Detector and
Divider cir-
cuitry
are functioning
properly
and the test equipment
can be
disconnected
from the
instrument.
INPUT
DC VOLTAGE
RANGE
NULL
SENS
.001
1
1 mv
.01
1
. Olv
. 1
1
. 1 v
1.0
1 TVM
10.0
10
TVM
100.0
100
TVM
1000.0
1000
TVM
Figure
4-5.
INPUT DC
VOLTAGE
VERSUS
RANGE-CHECKS
4-31.
KELVIN-VARLEY
DIVIDER
CHECKS
4-32.
The Kelvin-
Varley Divider
Checks require
the
use
of
a
dc voltage
calibrator.
The
Precalibration and
Troubleshooting
Checks of
paragraphs 4-21 through
4-30
must also have
been performed
to
ascertain that
the
associated
Instrument
circuitry
is functioning
correctly.
To perform
the checks,
apply
ac power
to the
instrument
and proceed
as follows:
a.
Place the Model 893
controls
as follows:
POWER
RANGE
NULL SENS
MODE
Readout dials
ON/LTNE
OPR
10
.001
+
10.
000
-
A4TP11
FULL-SCALE INPUT
Ft"".
;
5
u
SATURATION
CONDITION
4
1
:ffz
5
u —j
C
-
c
T...
u
A4TP13
A4TP15
Note:
HORIZONTAL
SWEEP
-
5
MSEC/CM
Figure
4-6.
NULL DETECTOR
WAVEFORMS
4-10
b. Apply
+10
volts
dc from
a dc voltage
calibrator
between
the
Model 893
LO
and HI
INPUT
terminals.
c.
Adjust
A2R33
for a zero-center
meter
indication
on the
Model
893
.
d. Apply the
dc
input
voltages
versus the
Readout
dial
settings
of Figure
4-7, observing
that the
Model
893 meter
indicates
zero-center
within the
major division
deviation.
INPUT
DC VOLTAGE
READOUT
DIAL
SETTINGS
MAJOR
DIVISION
DEVIATION
11.000
10. 99
X
±5.5
10.
000
10.00
0
null
reference
10. 00
9-99
X
±5. 0
9.999
9. 99
90
±4. 99
8. 888
8. 88
80 ±4,
44
7.777
'^77
20
±3. 88
6.666
6.66
^
±3.33
5.555
5.55
5(}
±2.77
4.444
4.44
40
±2.
22
3,333
3-33 30
±1,
66
2.222
2,22
20
±1.
n
1.111
1. 11 10
±
.55
1.000
1.00
Q
±
.5
Figure
4-7.
INPUT
VOLTAGES
VERSUS
READOUT
SETTINGS
-
CHECKS
4-33.
When
the results
of these
checks
agree with the
information
given, the
Kelvin-
Varley Divider
circuitry
is
functioning
properly
and the
test equipment
can be
disconnected
from the
instrument.
4-34.
RECORDER
OUTPUT
OPTION
CHECKS
4-35.
The
Recorder
Output
Option Checks
require
the
use
of a dc
differential
voltmeter.
Circuit
wave-
forms
are illustrated
in
Figure
4-8,
To
perform the
checks,
apply
ac power to the
instrument
and proceed
as
follows;
a. Connect the
inputs
of a dc differential
voltmeter
to
the
rear panel
HI and
LO
RECORDER OUTPUT
terminals and
rotate the
associated
level control
fully clockwise.
b. Connect
a shorting
jumper between
the
LO and
HI
INPUT
terminals
of the
Model
893
and
place the
controls
as follows:
POWER
RANGE
NULL SENS
MODE
Readout
dials
ON/LINE
OPR
1
.001
+
.00100
c. A
dc voltage
of
-0.
2 to
-0.
28 volts
dc should be
indicated
on the
dc differential
voltmeter
connected
to the
RECORDER
OUTPUT
terminals.
893A

Related product manuals