Wireless iFlex
Performance Tests
5
Required Equipment
The equipment in Table 2 is necessary for performance tests and calibration
adjustment.
Table 2. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator
4.5-digit resolution
DC Current Accuracy:
600 μA to 20 A ±0.25 %
Fluke 5522A Calibrator
(or equivalent)
Wired coil 50 turns 5500A/COIL
Test Probe for iFlex 2 mm to 4 mm Slim Reach Probe TP2, PN 650892
Test Lead
Test Lead with retractable sheath 6358,
PN1903307
Performance Tests
Warning
To prevent possible electrical shock, fire, or personal injury,
do not perform the performance test procedures unless the
Product is fully assembled.
The performance tests verify the full operation of the Product and measure the
accuracy of each function against Product specifications. If the Product fails a
part of the test, calibration adjustment and/or repair is necessary. See
“Calibration Adjustment”.
Test the Display
To verify that all segments of the display function:
1. With the Product off, push and hold .
2. Push while you keep pushed until all of the display segments are
shown. See Figure 1.
hby002.eps
Figure 1. All Segments of the Display
If segments of the display are missing, repair is necessary. See “Contact Fluke”.