Table 3. Performance Tests
Test Connection
Calibrator
Output Value
Res. Spec.
UUT Reading Limit
Low High
Single-Turn Loop
0 A,0 Hz 0.1
3.0 % ±5 digits
0 0.5
Single-Turn Loop 10 A,45 Hz 0.1 9.2 10.8
50-Turn Coil 12 A,45 Hz 0.1 581.5 618.5
50-Turn Coil 20 A,45 Hz 1 965 1035
Simulate 67.5 mV,45 Hz 1 2420 2580
Single-Turn Loop 10 A,500 Hz 0.1 9.2 10.8
Simulate 180 mV,500 Hz 0.1 581.5 618.5
Simulate 300 mV,500 Hz 1 965 1035
Simulate 750 mV,500 Hz 1 2420 2580
50-Turn Coil 45 Hz,0.4 A 0.1
0.5% ±5 digits
44.3 45.7
50-Turn Coil 500 Hz,0.325 A 0.1 497.0 503.0
0
/
+
5522A
HI
LO
TC
1000V
RM S
MAX
20V
RM S
MAX
20V
PK
MAX
1V PK
MAX
20V PK
MAX
NORMAL AUX
TRIG
OUT
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
150V PK
MAX
20V PK
MAX
a3001 FC
hca07.eps
Figure 2. Performance Test Connections for Simulated Voltages