789/787B
Users Manual
18
Measuring Electrical Parameters
The proper sequence for taking measurements is:
1. Plug the test leads into the appropriate jacks.
2. Set the rotary function switch to the desired function.
3. Touch the probes to the test points.
4. View the results on the LCD display.
Input Impedance
For the voltage measurement functions, input impedance
is 10 MΩ. See "Specifications" for more information.
Ranges
A measurement range determines the highest value and
resolution at which the meter can measure. Most meter
measurement functions have more than one range (see
"Specifications").
Make sure the correct range is selected:
If the range is too low, the display shows OL
(overload).
If the range is too high, the meter will not be
displaying its most precise measurement.
The meter normally selects the lowest range that will
measure the applied input signal (Auto Range showing on
the display). Press R to lock the range. Each time
R is pressed, the meter selects the next higher range.
At the highest range, it returns to the lowest range.
If the range is locked, the meter resumes auto ranging
when it is changed to another measurement function or
when R is pressed and held for 1 second.
Testing Diodes
To test a single diode:
1. Insert the red test lead into the Vjack and
black test lead into the COM jack.
2. Set the rotary function switch to
V.
3. Press J (Blue) so that the D symbol is on the
display.
4. Touch the red probe to the anode and the black
probe to the cathode (side with band or bands). The
meter should indicate the appropriate diode voltage
drop.
5. Reverse the probes. The meter displays OL,
indicating high impedance.
The diode is good if it passes the tests in steps 4
and 5.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ProcessMeter 787B and is the answer not in the manual?

Fluke ProcessMeter 787B Specifications

General IconGeneral
Battery life140 h
Battery sizeAA
Battery technologyAlkaline
Number of batteries supported4
ManualYes
Loop voltage1000 V
Drop proof up to1 m
Operating vibration2 G
Storage temperature (T-T)-40 - 60 °C
Operating temperature (T-T)-20 - 55 °C
Operating relative humidity (H-H)0 - 95 %
Weight and Dimensions IconWeight and Dimensions
Depth50 mm
Width100 mm
Height203 mm
Weight600 g