ProcessMeter™
Power-Up Options
27
Table 11. Power-Up Options
Option Button
Switch
Position
Default Display Action Taken
Current Span R All Retains setting -  or - 
Toggles between 0-20 mA and
4-20 mA range
Backlight Timeout K All Retains setting  / 
Enables/Disables the auto-off
on backlight after 2 minutes
Beeper G All Retains setting  /  Enables/Disables beeper
Auto Power-Off
Note: Auto power-off is
always disabled when
MIN MAX recording is on.
J
(Blue)
All Enabled

Disables the feature that turns
off the power after 30 minutes
of inactivity.
LCD segments H
VAC, mA,
Source,
Ramp,
Loop
Disabled All segments
Display HOLD (as long as
button pushed)
Firmware version H VDC Disabled ex: 
Display firmware version (as
long as button is pushed)
Model number H mVDC Disabled ex: 
Display model number (as long
as button is pushed)
Go to Calibration mode H Ω Disabled  Calibration mode starts

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ProcessMeter 787B and is the answer not in the manual?

Fluke ProcessMeter 787B Specifications

General IconGeneral
Battery life140 h
Battery sizeAA
Battery technologyAlkaline
Number of batteries supported4
ManualYes
Loop voltage1000 V
Drop proof up to1 m
Operating vibration2 G
Storage temperature (T-T)-40 - 60 °C
Operating temperature (T-T)-20 - 55 °C
Operating relative humidity (H-H)0 - 95 %
Weight and Dimensions IconWeight and Dimensions
Depth50 mm
Width100 mm
Height203 mm
Weight600 g