Characteristics
2.3 Dual Input Meter
2
2-7
Duty Cycle (DUTY)
Range 2% to 98%
Frequency Range for Continuous Autoset 15Hz (1Hz) to 30 MHz
Accuracy:
@1Hz to 1 MHz ±(0.5% +2 counts)
@1 MHz to 10 MHz ±(1.0% +2 counts)
@10 MHz to 40 MHz ±(2.5% +2 counts)
Resolution 0.1%
Pulse Width (PULSE)
Frequency Range for Continuous Autoset 15Hz (1Hz) to 30 MHz
Accuracy:
@1Hz to 1 MHz ±(0.5% +2 counts)
@1 MHz to 10 MHz ±(1.0% +2 counts)
@10 MHz to 40 MHz ±(2.5% +2 counts)
Full Scale reading 1000 counts
Amperes (AMP) with optional current probe
Ranges same as VDC, VAC, VAC+DC, or PEAK
Scale Factor 1 mV/A, 10 mV/A, 100 mV/A, and 1 V/A
Accuracy same as VDC, VAC, VAC+DC, or PEAK
(add current probe uncertainty)
Temperature (TEMP) with optional temperature probe
Range 200 °C/div (200 °F/div)
Scale Factor 1 mV/°C and 1 mV/°F
Accuracy as VDC (add temperature probe
uncertainty)
Decibel (dB)
0 dBV 1V
0 dBm (600 /50) 1 mW, referenced to 600 or 50
dB on VDC, VAC, or VAC+DC
Full Scale Reading 1000 counts
Crest Factor (CREST)
Range 1 to 10
Accuracy ±(5% +1 count)
Full Scale Reading 100 counts
Phase
Modes A to B, B to A
Range 0 to 359 degrees
Accuracy ±(1 degree +1 count)
Resolution 1 degree