123
Service Manual
2-10
Interface RS-232, optically isolated
To Printer supports Epson FX, LQ, and HP Deskjet
,
Laserjet
, and Postscript
Serial via PM9080 (optically isolated
RS232 adapter/cable, optional).
Parallel via PAC91 (optically isolated
print adapter cable, optional).
To PC Dump and load settings and data.
Serial via PM9080 (optically isolated
RS232 adapter/cable, optional), using
SW90W (FlukeView software for
Windows).
2.5 Environmental
Environmental MIL 28800E, Type 3, Class III, Style B
Temperature
Operating 0 to 50 °C (32 to 122 °F)
Storage -20 to 60 °C (-4 to 140 °F)
Humidity
Operating:
@0 to 10 °C (32 to 50 °F) noncondensing
@10 to 30 °C (50 to 86 °F) 95%
@30 to 40 °C (86 to 104 °F) 75%
@40 to 50 °C (104 to 122 °F) 45%
Storage:
@-20 to 60 °C (-4 to 140 °F) noncondensing
Altitude
Operating 4.5 km (15 000 feet)
Max. Input and Floating Voltage 600
Vrms up to 2 km, linearly derating to 400
Vrms @ 4.5 km
Storage 12 km (40 000 feet)
Vibration max. 3g
Shock max. 30g
Fungus Resistance MIL28800E, Class 3, 3.7.7 & 4.5.6.1
Salt Exposure MIL28800E, Class 3, 3.7.8.2 & 4.5.6.2.2.
Structural parts meet 48 hours 5% salt
solution test.
Electromagnetic Compatibility (EMC)
Emission EN 50081-1 (1992): EN55022 and
EN60555-2
Immunity EN 50082-2(1992): IEC1000-4-2, -3, -4, -5
(see also Section 2.8, Tables 2-1 to 2-3)
Enclosure Protection IP51, ref: IEC529