EasyManua.ls Logo

Fluke ScopeMeter 123 - Input a and B Base Line Jump Test

Fluke ScopeMeter 123
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
123
Service Manual
4-6
ST7968.CGM
Figure 4-2. Menu item selection
If an item is selected, it is marked by . Not selected items are marked by . If a
selected item is highlighted, an then
is pressed, the item remains selected.
You can also navigate through the menu using
. To conform the highlighted item
you must press
.
4.5.1 Input A and B Base Line Jump Test
Proceed as follows to check the Input A and Input B base line jump:
1. Short circuit the Input A and the Input B shielded banana sockets of the test tool.
Use the BB120 banana to BNC adapter, and a 50 (or lower) BNC termination.
2. Select the following test tool setup:
Turn Input B on (if not already on).
Press
to select auto ranging (AUTO in top of display).
(
toggles between AUTO and MANUAL ranging).
Press to open the SCOPE INPUTS menu.
Press
to open the SCOPE OPTIONS menu, and choose :
SCOPE MODE: NORMAL | WAVEFORM MODE: SMOOTH
3. Using toggle the time base between 10 ms/div and 5 ms/div.
(the time base ranging is set to manual now, the input sensitivity is still automatic; no
indication
AUTO or MANUAL is displayed).
After changing the time base wait some seconds until the trace has settled.
Observe the Input A trace, and check to see if it returns to the same position after
changing the time base. The allowed difference is ±0.04 division (= 1 pixel).
Observe the Input B trace for the same conditions.
4. Using
toggle the time base between 1 µs/div and 500 ns/div. After changing
the time base wait some seconds until the trace has settled.
Observe the Input A trace, and check to see if it is set to the same position after
changing the time base. The allowed difference is ±0.04 division (= 1 pixel).
Observe the Input B trace for the same conditions.

Table of Contents

Related product manuals