EasyManua.ls Logo

Fluke SMFT1000/PRO - Select PV Model; Paired with Irradiance Meter Only

Fluke SMFT1000/PRO
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SMFT-1000
Users Manual
22
Select PV Model
When the Irradiance Meter is not connected, no limits are available and no irradiance or
temperature data shows on the display.
To take a measurement:
1. Connect the test leads from the PV Analyzer to the solar panel.
Tip: Push
to see a connection diagram.
The V
OC
measurement shows on the display after you connect the test leads. The Pass/Fail
icons do not show in this configuration.
2. Push
to start the I
SC
measurement.
The V
OC
and I
SC
results show on the display.
3. Push
to save the results to memory.
A confirmation message with ID number shows on the display and then returns to the test
screen.
Paired with Irradiance Meter Only
When the Irradiance Meter is connected and a PV model is not selected, no limits are available.
The irradiance and temperature data from the Irradiance Meter shows on the display.
To take a measurement:
1. Connect the test leads from the PV Analyzer to the solar panel. The V
OC
measurement
automatically shows on the display.
Tip: Push
to see a connection diagram.
The V
OC
measurement shows on the display after you connect the test leads. Irr & Tcell
data from the Irradiance Meter shows on the display. The Pass/Fail icons do not show in
this configuration.
2. Push
to start the I
SC
measurement.
The V
OC
and I
SC
results show on the display.
3. Push
to save the results to memory.
A confirmation message with ID number shows on the display and then returns to the test
screen.