EasyManuals Logo

Fluke T3 User Manual

Fluke T3
14 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
Tester
Parts and Service
3
Parts and Service
The tester is warranted to be free from defects in material and workmanship for 1 year, while
under normal use. Parts and repairs are warranted for 90 days. For the complete warranty
statement, refer to the T3 Tester Instruction Sheet.
To order parts, or for warranty service, contact Fluke as follows:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke’s Web site at www.fluke.com.
Specifications
Display Accuracy The LED for each range turns on by 95 % of the nominal range value.
Maximum Voltage
Between any Terminal
and Earth Ground
1000 V dc; 1000 V ac rms (sine wave), Overvoltage Category III; Tester
meets requirements for CAT IV 600 V, allowing use on outdoor/direct burial
wiring or utility-side measurements.
Input Impedance ~750 k
Temperature Operating: -10 °C to +50 °C (14 °F to 122 °F) Storage: -30 °C to +60 °C
(-22 °F to +140 °F)
Altitude Operating: 3000 m (9843 ft); Storage: 10,000 m (32808 ft)
Relative Humidity 0 °C to 30 °C (32 °F to 86 °F): 90 %; 30 °C to 40 °C (86 °F to 104 °F): 75 %;
40 °C to 50 °C
(104 °F to 122 °F): 45 %
Battery Type and Life AA (2); 250 hours with NEDA 15F or IEC R6
Shock, Vibration 1 m drop at 15 °C to 35 °C (59 °F to 95 °F). Sinusoidal vibration per MIL-
PRF-28800F for a Class 2 instrument (5 Hz to 55 Hz, 3 g maximum)
Environmental Seal IP 52 per IEC 529, no vacuum applied
Safety This tester complies with IEC 1010-1 to 1000V OVERVOLTAGE Category III,
Pollution Degree 2, and with IEC 664-1 to 600 V OVERVOLTAGE Category
IV, Pollution Degree 2.*
*OVERVOLTAGE (Installation) Categories refer to the level of Impulse
Withstand Voltage protection provided at the specified Pollution Degree.
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed
installations. Examples include switchgear and polyphase motors.
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the
installation. Examples include electricity meter and primary over-current
protection equipment.
EMC Regulations EN61326
Certifications
, , Listed 950Z N10140, VDE (Pending),