REF4P, REF5P and REF5BBP-T Undercounter Refrigerators 27
HEALTHCARE
TM
Parameter Default Value Range Description
Door Heater
Door Heater Yes Yes, No Controls door heater output
Door Heater Off
(mins)
2 0-100 Off time for door heater if dht=on
Door Heater On
(mins)
3 0-100 On time for door heater if dht=on
Error control
Probe Error BPr NO, BPr, R1 relay, R2 relay Action to take when probe error detected
Control On Time
(mins)
3 minutes 0.0 - 120.0 minutes Compressor on time when control probe error
(minutes)
Control Off Time
(mins)
10 minutes 0.0 - 120.0 minutes Compressor off time when control probe error
Fail Safe Timer
(mins)
0 minutes 0 - 180 minutes Minimal time the suction or discharge fail safe
routines must remain off when triggered.
Calibration
Calibrate probes Calibrate probes through a second menu
Viewable
Cut In View cut in temperature
Cut Out View cut out temperature
Probe 1 temp View temperature of probe 1
Probe 3 temp View temperature of probe 3
Power On Time
(hrs)
View cumlitive hours that the unit was in service
Compressor
cycles
The number of compressor starts
Compressor run
time(hrs)
Veiw cumulative hours that the compressor was
energized
Door open Cycles The number of door opening
Door open
time(hrs)
View cumulative hours that the door was open
Factory Reset
Reset No No, Yes Reset all parameters to factory reset