14. ELECTROMAGNETIC COMPATIBILITY
14.1. ELECTROMAGNETIC EMISSION
The FONA XPan 3D Plus system is suitable for use in the electromagnetic environment specified below. The customer
or the user of FONA XPan 3D Plus system should assure that it is used in such an environment.
Electromagnetic environment
Radiated and conducted RF emissions
CISPR 11
This FONA XPan 3D Plus uses RF energy only for its
internal function. Therefore. the RF emission is very
low and not likely to cause any interference in nearby
electronic equipment.
This FONA XPan 3D Plus is suitable for use in all
establishments. including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes
Harmonic Emissions EN 61000-3-2
Voltage fluctuations/ flicker emissions
61000-3-3
14.2. ELECTROMAGNETIC IMMUNITY
The FONA XPan 3D Plus system is intended for use in the electromagnetic environment specified below. The
customer or the user of the FONA XPan 3D Plus system should assure that it is used in such an environment.
Electromagnetic environment
Electrostatic discharge (ESD)
IEC 61000-4-2
Floors should be wood. concrete
or ceramic tile. If floors are
covered with synthetic material.
the relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/output
lines > 3 m
Mains power quality should be
that of a typical commercial or
hospital environment.
1 kV differential mode
2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips. short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
0% Un for 0.5 cycles
40 % Un for 5 cycles
70 % Un for 25 cycles
0 % Un for 5 s
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the FONA XPan 3D Plus
requires continued operation
during power mains
interruptions. it is recommended
that the FONA XPan 3D Plus be
powered from an uninterruptible
power supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.