49
Error Messages
MESSAGE WHAT IT MEANS WHAT TO DO
Err 1
Err 2
Inflation or pressure error.
Refit cuff tightly and correctly. Relax
and repeat the measurement. If the
error still remains, please contact local
customer service for help.
Err 3
Cuff pressure is too high.
Err 4
Err 5
No pulse rate detected or
the pulse rate is too low.
Err 6
Systolic pressure is
outside the systemʼs
measurement range
System measurement range is 60
mmHg – 250 mmHg. Review the
instructions and repeat the test. If the
error still remains, please contact your
healthcare professional for help.
Err 8
Forced to stop the blood
pressure measurement
after re-pump for 2 times.
Turn off the device and start over.
Follow the instructions and try again.
Err 15
Data corrupted.
Follow the instructions and try again.
If problem persists, contact the local
customer service for assistance.
Appears when the
batteries cannot provide
enough power for a test.
Replace or recharge the batteries
immediately.
E-U
Strip has been used. Repeat the test with a new strip.
E-t HIGH
E-t LOW
Environmental
temperature is outside
the systemʼs operational
range.
System operational range is 50
°
F to
104
°
F (10
°
C to 40
°
C). Repeat the
test after the meter and test strip have
reached the above temperature.
E-d
Problem in operation.
Review the instructions and repeat the
test with a new strip.