EasyManua.ls Logo

Fora G20 - Specifications

Fora G20
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
51
SPECIFICATIONS
Model No.: TD-4251
Dimension & Weight: 86.9mm(L) x 51mm(W) x 15.8mm(H), 42.2g
Power source: One CR2032
Memory: 450 measurement results with date and time
Auto electrode inserting detection
Auto sample loading detection
Auto reaction time count-down
Auto turn-off after 3 minutes without action
Operating condition: 50°F ~104°F(10°C~40°C), below 85% R.H.(noncondensing)
Meter storage/Transportation condition: -4°F~140°F(-20°C~60°C), below 95% R.H.
Measurement Units: Fix mg/dL
Measurement Range: 20~600mg/dL (1.1~33.3mmol/L)
The device has been certied to meet the electrical and safety requirements of:
IEC 60601-1, EN 60601-1, IEC 61010-1, EN 61010-1, EN 61010-2-101, EN
60601-1-2, EN 61326.
Performance characteristics
Accuracy
Within ±15 mg/dL (0.83 mmol/L) at glucose concentration < 75 mg/dL
(4.2 mmol/L) and within ±20% at glucose concentration ≥ 75 mg/dL (4.2
mmol/L).
Precision
CVs (%) of intermediate precision and repeatability are less than 5%.
The product has been tested to meet the requirements of ISO 15197.

Table of Contents

Other manuals for Fora G20

Related product manuals