18
2. If the test does not start after applying the
sample:
3. If the control solution testing result is out of
range.
Model No.: TD-4251
Dimension & Weight: 86.9mm(L) x 51mm(W) x
15.8mm(H), 42.2g
Power Source: one CR2032 lithium battery
Display: LCD
Memory: 448 measurement results with
respective date and time
External output: RS232 PC interface
Auto electrode insertion detection
Auto sample loading detection
Auto reaction time count-down
Auto switch-off after 3 minutes without action
Temperature warning
Operating Condition: 50°F to 104°F (10°C to
40°C), below 85% R.H. (non-condensing)
Storage/Transportation Conditions:
Meter: -4°F to 140°F (-20°C to 60°C), below 95%
R.H.
Test strip: 35.6°F to 89.6°F (2°C to 32°C), below
85% R.H. Do not freeze.
Measurement Units: mg/dL
Measurement Range: 20 to 600mg/dL (1.1 to
33.3mmol/L)
This device has been tested to meet the electrical
and safety requirements of: IEC/EN 61010-1, IEC/
EN 61010-2-101, EN 61326-1, IEC/EN 61326-2-2.
Distributed by: Manufactured by:
SPECIFICATIONS
9247 Research Drive · Irvine, CA
Toll Free: (888) 425-1149
(8:00am-5:00pm PST, Mon.-Fri.)
www.linksmed.com
POSSIBLE CAUSE
WHAT TO DO
POSSIBLE CAUSE
WHAT TO DO
.
.
.
Insuffi cient blood sample.
Defective test strip.
Sample applied after
automatic switch-off
(3 minutes after last
user action).
Defective meter.
Re-test using a new test
strip with larger volume of
blood.
Re-test with a new test
strip.
Re-test with a new test
strip. Apply sample only
when " " appears
on the display.
Please contact Customer
Service at (888) 425-1149.
Error in performing the test.
Control solution vial was
porly shaken.
Expired or contaminated
control solution.
Control solution that is too
warm or too cold.
Defective test strip.
Meter malfunction.
Read the instructions
thoroughly and repeat the
test again.
Shake the control solution
vial and repeat the test
again.
Check the expiration date
or the discard date of the
control solution.
Control solution, meter, and
test strips should come to
room temperature (68°F to
77°F(20°C to 25°C) before
testing.
Re-test with a new test strip.
Please contact Customer
Service at (888) 425-1149.
ForaCare Inc.
893 Patriot Dr., Suite D,
Moorpark, CA 93021