Maximum power consumption
110V (Watts) (BoSS System)
1600 1700 2400 1700 2400 1700
Power consumption* 110V, 60
Hz (Watts)
1000 1100 1400 1100 1400 1100
Minimum power of cooling of
the room containing the freezer
in continuously (W)
850 950 1200 950 1200 950
Acoustic pressure measured in
dB(A) (BoSS System)
76
Acoustic pressure measured in
dB(A) (in regulation mode)
56
OUTSIDE DIMENSIONS
Height (mm) 1280 1280 1640 1640 2000 2000
Width (mm) 875 875 875 875 875 875
Depth (mm) 970 970 970 970 970 970
Clearance required at the rear of
the unit (mm)
200
200 200 200 200 200
Maximum dimension with door
open (mm)
1750
1750 1750 1750 1750 1750
Maximum width with door open
(mm)
1150
1150 1150 1150 1150 1150
Maximum door opening angle 110° 110° 110° 11 110° 110°
GROSS INTERNAL
DIMENSIONS EXCLUDING
RACKS
Height (mm) 716 716 1076 1076 1436 1436
Width (mm) 630 630 630 630 630 630
Depth (mm) 752 752 752 752 752 752
Notes
* Power consumed in regulation mode, all parts activated and stabilized, at 23 ° C in atmosphere and with empty cabinet.
On an equivalent product, hydrocarbon (HC) gases allow a reduction in power consumption up to 25%, as the need of cooling.
Refrigeration system:
FREEZER MODEL BM340 BM515 BM690
Standard
cooling
Natural
cooling
Standard
cooling
Natural
cooling
Standard
cooling
Natural
cooling
TWO-STAGE -86°C MODELS
Stage 1 hermetic compressor
power (Watt)
580
380 780 580 780 580
Stage 2 hermetic compressor
power (Watt)
380
380 580 580 580 580