Section 6 Parameter Settings
59 FT742-DM (RS485) Sensors ā User Manual
6.4.22 UC.2: Clear User Calibration Table Record
$<listenerID>,UC<erase>*<checksum><cr><lf>
$aa,UCCLEAR*hh<cr><lf>
Erases both Flash and RAM copies of user calibration table
Example 1
Erase the user calibration tables and verify
Query user calibration table status
$WI,UC=00,D,0000,0000*71<cr><lf>
Use the UCCLEAR command to erase the RAM and saved FLASH copies of the user
calibration table. A UCCLEAR is performed before a new user calibration table is loaded
(see Section 6.4.23).
The user calibration table label is also cleared to 32 ASCII spaces when the UCCLEAR
command is sent (see Section 6.4.25).