44
Normal Mode
Setup Mode
Ad0
ADG
CJTL
CJG
REF
SR
4 seconds
4 seconds
RESET
RESET
RESET
RESET
RESET
RESET
RESET
Step 1
4 seconds
Step 2
4 seconds
Step 3
4 seconds
Step 4
4 seconds
Step 5
4 seconds
Step 6
4 seconds
Figure 5-1
Flow Chart for Manual
Calibraton
Figure 5-1
Flow
Chart for Manual
Calibraton
Step1: Calibrate of A to D converter.
Short terminal 4 and 5, then press for at least 4 seconds.
The display will blink a moment. If the display didn't blink,
then the calibration has failed.
Zero
UM L91-Rev 8