EasyManua.ls Logo

Future Design FDC-L91 - Page 45

Default Icon
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Step 2: Calibrate of A to D converter.
Send a span signal to terminal 4 and 5 with correct polarity.
The span signal is 60 mV for thermocouple input, 1V for
0-1V input, 10V for 0-10V input and 20mA for 0-20 mA input.
Press for at least 4 seconds. The display will blink a
moment. If the display didn't blink, then the calibration failed.
Gain
Step 3: Calibrate of .
Setup the equipment according to the following diagram
for calibrating the cold junction compensation. Note that a
K type thermocouple must be used.
offset cold junction
offset cold junction
4
5
Set the calibrator to be configured as K type thermocouple
output. Calibrator must have an internal compensation.
Send a 0.00 C signal to the unit under calibration.
Calibrator
K-TC
K+
K
NOTE: The unit under calibration is powered in a still-air room
at a temperature . Allow at least 20 minutes for
warming up.
25 +/-3 C
NOTE: The unit under calibration is powered in a still-air room
at
a temperature 25 +/-3 C. Allow at least 20 minutes for
warming
up.
Figure 5-2
Cold Junction
Calibration Setup
Figure 5-2
Cold Junction
Calibration
Setup
45
L91
Stay at least 20 minutes in still-
air room room temperature
25 +/- 3 C
UM L91-Rev 8
With CJTL on the display adjust the value to 0.00 reading.
Once adjusted, Press for at least 4 seconds. The display will
blink a moment. If the display didn't blink, then the calibration
failed.
The L91 being calibrated for Cold Junction Compensation MUST
be programmed for K t/c input, Celsius display to performing
the CJTL calibration.
prior

Related product manuals