GBA ST1C Operation Manual
Page 11 of 30
Astrosys International Ltd
GBA ST1C Operations manual V1_4
4.1.2 Configuration Procedure – Disable Note
Procedure:
1. Push button x 2, the Status LED will flash red rapidly.
2. Insert a sample note of the specific denomination required to be disabled.
3. The GBA ST1C will scan the note in and then return it.
4. Disable complete. The Status LED will return to solid green.
4.1.3 Configuration Procedure – Programming Card
Equipment Required;
Program Card. (PC software is available to print these cards. Please consult your local sales office)
Procedure;
1. Push button x 3, the Status LED will flash orange rapidly.
2. Insert the Program Card into the reader.
3. The GBA ST1C will scan the Program Card in and then return it.
4. Configuration is complete.
5. If programming has not been successful, the Status LED will repeatedly flash a 4 colour
error code until the push button is pressed to cancel.
6. If successful, the Status LED will return to steady green or orange.
Note: Please ensure the programming card has been printed on a high quality printer with good
solid black lines on the card.
4.1.4 Configuration Procedure – Re-Calibration
Although calibration of the GBA ST1C is not a necessity it is recommended periodically for
preventative maintenance, and also after any change to the software configuration of the unit.
Equipment Required;
Calibration Papers:
Synthetic White Paper - 2 chamfered corners, Part Number ASY-W-10268 for units with US
or Canadian note guides, Part Number ASY-W-10311 for units with wider note guides.
Black calibration card
Procedure;
1. Push button x 3, Status LED will rapidly flash orange.
2. Insert the white calibration paper into the validator. It should be presented in the centre of
the note path and square to the direction of travel.
3. The ST1C unit will draw-in the paper all the way into the cassette; hold it momentarily
before returning it in several short steps.
4. Remove the calibration paper from the note entry point.
5. Insert the Black calibration card.
6. The ST1C unit will draw-in the card and return it immediately in one smooth action
7. Calibration is complete
8. If calibration has not been successful, the Status LED will repeatedly flash a 4 colour error
code until the push button is pressed to cancel.
9. If successful, the Status LED will return to steady green/orange.
Note: Please ensure the ST1C calibration paper has 2 chamfered corners.