APPENDIX A: EMC INFORMATION
If you have any questions regarding the guidance and declarations listed below, please
contact your home care provider.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions: This device is
intended for use in the electromagnetic environment specified below. The user of this
device should make sure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC
ENVIRONMENTGUIDANCE
RF emissions CISPR 11 Group 1 This device uses RF energy
only for its internal function.
Therefore, its RF emissions
are very low and are not
likely to cause any interfe-
rence in nearby electronic
equipment.
RF emissions CISPR 11 Class B The device is suitable for
use in all establishments,
including domestic estab-
lishments and those directly
connected to the public
low-voltage power supply
network.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/Flicker Complies
emissions IEC 61000-3-3
Emission of Radio Frequen-
cy Energy
RTCA/DO-160G Section 21
Category M This device is suitable for
use onboard commercial air-
planes inside the passenger
cabin.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity: This device is
intended for use in the electromagnetic environment specified below. The user of this
device should make sure that it is used in such an environment.
IMMUNITY TEST IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT
GUIDANCE
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±15kV air
±8kV contact
±15kV air
±8kV contact
Floors should be wood,
concrete or ceramic
tile. If floors are co-
vered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for input-output
lines
±2 kV for supply
mains
±1 kV for input/out-
put lines
Mains power quality
should be that of a ty-
pical home or hospital
environment.
Surge
IEC 61000-4-5
±1 kV dierential
mode
±2 kV common
mode
±1 kV dierential
mode
±2 kV common
mode
Mains power quality
should be that of a ty-
pical home or hospital
environment.