Planned maintenance
3-31
Document no. 2044677-001
2. Service reset
Check the Service Reset. Disconnect the power cord and remove the batteries to reset the
monitor. Switch the monitor back on and check that the monitor performs a Cold Start, i.e.
all trend information is cleared.
NOTE: Cold start have three ways:
Disconnect all power source and reset the monitor
When the monitor power on, hold on the
Power hardkey 15 seconds
Turn off the monitor, waiting for 15 minutes
"
3. Watchdog
Enter the Set/Test service menu and perform Watchdog
Check that the monitor restarts.
"
4. Service Log reset
Enter the Service Log service menu.
Clear the content of the Service Log by selecting Reset Log from the menu.
"
5. Electrical safety check
Perform an Electrical safety check and a leakage current test. Check that the monitor and
all connected units function normally after the performed test.
"
6. Final cleaning
Switch off the monitor and perform final cleaning.
Fill in all necessary documents.
"

Table of Contents

Other manuals for GE b30

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE b30 and is the answer not in the manual?

GE b30 Specifications

General IconGeneral
Modelb30
ManufacturerGE
Coil VoltageN/A (Solid State Relay)
Coil PowerN/A (Solid State Relay)
Mechanical LifeN/A (Solid State Relay)
Electrical LifeN/A (Solid State Relay)

Related product manuals