B30 Patient Monitor
5-4
Document no. 2044677-001
2.2 Network
2.2.1 Network
Network: For setting the monitor's network mode:
S5/Unity
NOTE: Only in Chinese language, Unity mode is
available.
2.2.2 Dri Config
The DRI Level: For setting the monitor's network
communication. The network communication needs
match to the iCentral’s configuration
Transfer Mode: For choosing the transfer mode:
DRI/ETH or DRI/UDP
Default SLP IP: For R&D purpose only. (DRI/ETH is not
in use)
Forced Central IP: For setting IP address, which
according to the iCentral’s configuration. (DRI/ETH is
not in use, it’s can capture the IP automatically)
Forced Central Port: For setting port address, which
according to the iCentral’s configuration. (DRI/ETH is
not in use, it’s can capture the port automatically)
Debug server IP: For R&D purpose only.
Software download: For setting software download
capability.
Save Changes: Select "Save Changes" to take in use
changes made in this page.