B30 Patient Monitor
7-2
Document no. 2044677-001
1.3 Batteries
Number of batteries Support up to 2
Battery type Exchangeable Lithium-Ion
Voltage 11.1V (nominal)
Capacity 3.5 Ah
Capacity indicator Fuel gauge on monitor screen and on battery pack
Run time Up to 2:15 hours per battery pack
Charge time 2 hours per battery pack
Battery life 300 cycles to 50% capacity
Conditioning Manual
1.4 Network
Connector type RJ-45
Network type 10/100 Base-T
Isolation 1500 V, 60/50Hz, 1 min.
Network ID Network ID key
Communication protocol DRI 01, 02 and 03
Network Connectivity D-O Central and iCentral
1.5 Analog Outputs, direct ECG and synchronization
ECG Direct ECG delay 15 ms (max.)
Gain ECG (in)/ECG(out) 1 mV/1V
Pacer 5 V and 2 ms pulse
InvBP InvBP Output ‘Art’ or P1
InvBP Delay 35ms (max.)
Gain InvBP 10mV/mmHg
Defibrillation Synchronization State 0: 0 - 0.5V; state 1: 3.5 - 5V
Delay (max): < 35 ms from R peak
5 V and 10 ms pulse
1.6 Digital outputs
Nurse call Low State: 0 - 0.5V; High State: 3.5 - 5V
Generated on red and yellow alarms
Serial Output Asynchronous
Serial data interface, uses RS-232 standard