EasyManuals Logo

GE b30 Technical Reference Manual

GE b30
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #194 background imageLoading...
Page #194 background image
B30 Patient Monitor
7-2
Document no. 2044677-001
1.3 Batteries
Number of batteries Support up to 2
Battery type Exchangeable Lithium-Ion
Voltage 11.1V (nominal)
Capacity 3.5 Ah
Capacity indicator Fuel gauge on monitor screen and on battery pack
Run time Up to 2:15 hours per battery pack
Charge time 2 hours per battery pack
Battery life 300 cycles to 50% capacity
Conditioning Manual
1.4 Network
Connector type RJ-45
Network type 10/100 Base-T
Isolation 1500 V, 60/50Hz, 1 min.
Network ID Network ID key
Communication protocol DRI 01, 02 and 03
Network Connectivity D-O Central and iCentral
1.5 Analog Outputs, direct ECG and synchronization
ECG Direct ECG delay 15 ms (max.)
Gain ECG (in)/ECG(out) 1 mV/1V
Pacer 5 V and 2 ms pulse
InvBP InvBP Output ‘Art’ or P1
InvBP Delay 35ms (max.)
Gain InvBP 10mV/mmHg
Defibrillation Synchronization State 0: 0 - 0.5V; state 1: 3.5 - 5V
Delay (max): < 35 ms from R peak
5 V and 10 ms pulse
1.6 Digital outputs
Nurse call Low State: 0 - 0.5V; High State: 3.5 - 5V
Generated on red and yellow alarms
Serial Output Asynchronous
Serial data interface, uses RS-232 standard

Table of Contents

Other manuals for GE b30

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE b30 and is the answer not in the manual?

GE b30 Specifications

General IconGeneral
Modelb30
ManufacturerGE
Coil VoltageN/A (Solid State Relay)
Coil PowerN/A (Solid State Relay)
Mechanical LifeN/A (Solid State Relay)
Electrical LifeN/A (Solid State Relay)

Related product manuals